Absolute and
gauge pressure
Cerabar PMP71

Cerabar PMP71 aluminum housing ©Endress+Hauser
Cerabar PMP71 stainless steel housing ©Endress+Hauser
Cerabar PMP71 sperate housing ©Endress+Hauser

Pressure measuring principle

Cerabar PMP71 application ©Endress+Hauser
Cerabar PMP71 application ©Endress+Hauser
Cerabar PMP71 with manifolds ©Endress+Hauser

Digital pressure transmitter with welded metal sensor for measurement in gases, steam or liquids

The Cerabar PMP71 digital pressure transmitter with metallic membrane is typically used in process applications for pressure, level, volume or mass measurement in liquids or gases. PMP71 is designed for high pressure applications up to 700bar. Quick Setup with freely adjustable measuring range without pressure specification. Designed according to IEC 61508 for use in SIL2/3 safety systems and available with MID Parts Certificate suitable for custody transfer applications.

  • Edut

    • Best accuracy, reproducibility and long-term stability

    • Highest safety due to gas tight feedthrough with capabilities up to SIL2/3, certified to IEC 61508

    • Easy menu-guided commissioning via local display, 4 to 20mA with HART, PROFIBUS PA, FOUNDATION Fieldbus

    • HistoROM data management concept for fast and easy commissioning, maintenance and diagnostics

    • ...

  • Käyttöalue

    Two chamber aluminum or stainless steel (316L) housing.

    • Process connections: Threads and flange connections

    • Process temperature: -40 to +125°C (-40 to +257°F)

    • Measuring ranges: -1/0 up to 700bar (-15/0 up to +10,500psi)

    • Accuracy: ±0.05%, "Platinum" ±0.025%

    • International explosion...

Ominaisuudet ja tiedot

Takaisin

Yksityiskohtaiset tiedot

  • Mittausperiaate

    Absolute and gauge pressure

  • Characteristic

    Digital transmitter with piezoresistive sensor and welded metallic membrane
    Modular transmitter
    Long term stability
    Enhanced safety via self diagnostic functionsSecondary process barrier

  • Supply voltage

    4...20 mA HART

    10,5...45V DC (Non Ex):

    Ex ia: 10,5...30V DC

    PROFIBUS PA:

    9...32 V DC (Non Ex)

    FOUNDATION Fieldbus:

    9...32 V DC (Non Ex)

  • Reference Accuracy

    Standard: 0.05%

    Platinum: up to 0.025%

  • Long term stability

    0.05 % of URL/ year

    0.07 % of URL/ 5 years

    0.1 % of URL/ 10 years

  • Process temperature

    -40°C...125°C
    (-40°F...257°F)

  • Ambient temperature

    -50°C...85°C
    (-58°F...185°F)

  • Measuring cell

    100 mbar...700 bar
    (1.5 psi...10.500 psi)

    relative/ absolute

  • Smallest calibratable span

    5 mbar (0.075 psi)

  • Vacuum resistance

    10 mbar (0.15 psi)

  • Max. Turn down

    100:1

  • Max. overpressure limit

    1050 bar (15.750 psi)

  • Process connection

    Thread:

    G1/2...G2, R1/2, MNPT1/2...MNPT2, R1/2

    Flange:

    DN25...DN80,

    ASME 1"...4",

    JIS 10K...20K

  • Material process membrane

    316L, AlloyC,

    Rhodium> Gold

  • Material gasket

    None, measuring cell welded

  • Fill fluid

    Silicone oil,

    Inert oil

  • Material housing

    Die-cast aluminum,

    AISI 316L

  • Communication

    4...20 mA HART

    1- 5V DC Low Power

    PROFIBUS PA

    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC

  • Safety approvals

    Overfill prevention WHG

    SIL

  • Design approvals

    EN10204-3.1

    NACE MR0175

  • Marine approvals

    GL/ABS

  • Drinking water approvals

    NSF

  • Specialities

    Diagnostic functions

  • Mittausperiaate

    Absolute and gauge pressure

  • Characteristic / Application

    Digital transmitter with piezoresistive sensor and welded metallic membrane
    Modular transmitter
    Long term stability
    Enhanced safety via self diagnostic functions
    Secondary process barrier

  • Specialities

    diagnostic functionalities

    different languages in software

  • Supply / Communication

    4...20mA HART:
    10,5...45V DC
    Ex ia: 10,5...30V DC
    PROFIBUS PA /
    FOUNDATION Fieldbus:
    9...32V DC

  • Accuracy

    0,075%
    "PLATINUM" 0,05%

  • Long term stability

    0,05% of URL/year

  • Ambient temperature

    -50°C...85°C
    (-58°F...185°F)

  • Process temperature

    -40°C...125°C
    (-40°F...257°F)

  • Process pressure absolute / max. overpressure limit

    1050bar (15,750psi)

  • Pressure measuring range

    100mbar...700bar
    (1.5psi...10.500psi)

  • Main wetted parts

    Alloy C276
    316L
    Rhodium

  • Process connection

    Threads
    Flanges (DIN, ASME, ...) with flush membrane

  • Max. measurement distance

    7000m (22.966ft) H2O

  • Communication

    4...20 mA HART

    1- 5V DC Low Power

    PROFIBUS PA

    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC

  • Safety approvals

    Overfill prevention WHG

    SIL

  • Design approvals

    EN 10204-3.1

    NACE MR0175, MR0103

  • Marine approval

    GL/ ABS

  • Drinking water approvals

    NSF

  • Options

    HistoROM/M-Dat
    4-line digital display
    SS- or Aluminiumhousing
    Separate housing

  • Application limits

    Measuring cell: Metal
    welded
    If pressurized, possibly
    use differential pressure
    meas-urement with two
    pressure transmitters
    (electronic dp). Observe
    ratio head pressure :
    hydrostatic pressure

Asiakirjat / käsikirjat / ohjelmisto

Tarvikkeet / varaosat

Nro
Tilauskoodi
Erittely
  • 12
    • Nro 12 Tilauskoodi 52020440

      Mounting set housing T14/T15 / sensor

      Consists of: 2x O-ring 45.69x2.62 EPDM. Retaining ring, plastics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 12 Tilauskoodi 71020596

      Mounting set housing T17 / sensor

      Consists of: 2x O-ring 45.62x69x2.62 EPDM. 1x profile gasket 51.4x57.2x1.6 EPDM. Retaining ring, plastics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

  • 13
    • Nro 13 Tilauskoodi 52022418

      Push button housing, PA/FF

      Consists of: push button, cover + screws. Usage: external operation, FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 13 Tilauskoodi 52022419

      Push button housing, HART, version 1

      Consists of: push button, cover + screws. Colour light grey. Usage: external operation, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 13 Tilauskoodi 52024110

      Push button housing, HART, version 2

      Consists of: push button, cover + screws. Colour dark grey. Usage: external operation, FMB70, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75.

  • 15
    • Tilauskoodi 71111056

      Housing filter VA,EPDM,F15/31/T14/17,1pc

      Consists of: vent screw 304. Pressure compensation filter. O-ring EPDM, protection lid. Usage: housing T14, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. Housing F31, FMB50/FMB51/FMB52/FMB53, PMC51, PMP51/PMP55. Housing F15 w/o B1,BB,BD,BF,GA,ID,I1, FMB50/FMB51/FMB52/FMB53, PMC51, PMP51/PMP55.

    • Tilauskoodi 71219639

      Housing filter VA, VMQ, F15, T17, 1 pc.

      Consists of: vent screw 304. Pressure compensation filter. O-ring VMQ, protection lid. Usage: housing T17, FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/ PMP75. Housing F15 bei B1,BB,BD,BF,GA,ID,I1, FMB50/FMB51/FMB52/FMB53, PMC51, PMP51/PMP55.

  • 20
    • Nro 20 Tilauskoodi 52020432

      Cover T14/T15 Alu, gasket

      Material: aluminium, coated. O-ring EPDM, FDA. Not for EEx d/XP. Usage: FMB70, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 52020433

      Cover T14/T15 Alu, Ex d, gasket

      Material: aluminium, coated. O-ring EPDM, FDA. EEx d/XP. Usage: FMB70, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 52028310

      Cover T14 316L, Ex d, EPDM

      Material: 316L. O-ring EPDM, FDA. Individual tested, EEx d/XP. Usage: FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 71002774

      Cover T17 316L, gasket

      Material: 316L. Flat gasket EPDM, centering ring, screw. Not for EEx d/XP. Usage: FMB70, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 71002810

      Cover T17 316L, sight glass, gasket

      Material: 316L. Sight glass: mineral glass. Flat gasket EPDM, centering ring, screw. For approval 2,3,4,7,8,Q,W. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. For approval BC,BD,IB. Usage: FMD71/FMD72.

    • Nro 20 Tilauskoodi 71002811

      Cover T17 316L, sight glass PC, gasket

      Material: 316L. Sight glass: Polycarbonate PC. Flat gasket EPDM, centering ring, screw. For approval A,E,F,H,I,R,S,U,1,6. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. For approval AA,BA,CA,CB,CC,CD,FA,FB, FD,IA. Usage: FMD71/FMD72.

    • Nro 20 Tilauskoodi 71206965

      Cover T14 316L, Ex d, FVMQ

      Material: 316L. O-ring FVMQ. Individual tested, EEx d/XP. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP72/PMP75.

    • Nro 20 Tilauskoodi 71206968

      Cover T14 316L, Ex d, sight glass, FVMQ

      Material: 316L. Sight glass: mineral glass,coated. O-ring FVMQ. Individual tested, EEx d/XP. Usage: FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

  • 21
  • 20
    • Nro 20 Tilauskoodi 71245247

      Cover T14/T15 Alu, sight glass, gasket

      Material: aluminium, coated. Sight glass: mineral glass, coated. O-ring EPDM, FDA. Not for EEx d/XP. Usage: FMB70, FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 71245249

      Cover T14/T15 Alu,Exd,sight glass,gasket

      Material: aluminium. Sight glass mineral glass, coated. O-ring EPDM, FDA. Individual tested, EEx d/XP. Usage: FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 20 Tilauskoodi 71245250

      Cover T14 316L, Ex d, sight glass, EPDM

      Material: 316L. Sight glass: mineral glass, coated. O-ring EPDM, FDA. Individual tested, EEx d/XP. Usage: FMD71/FMD72/FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

  • 21
    • Tilauskoodi 71290825

      Ground terminal T14/T15/T17, F30/F31

      Consists of: spring washer, screw, earth handle, clamp. Usage: housing T14/T15/T17, FMB70, FMD71/FMD72/ FMD77/FMD78, PMC71, PMP71/PMP75, PMD70/PMD75. Housing F30/F31, FMB50/FMB51/FMB52/ FMB53, PMC51, PMD55, PMP51/PMP55.

  • 25
  • 30
    • Nro 30 Tilauskoodi 52024111

      Electronics HART Ex, ext. p. b., v2.0

      4-20mA, version 2.0. Push button on housing. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 52024400

      Electronics HART Ex, v2.0

      4-20mA, version 2.0. Push button on electronics. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71026358

      Electronics HART Ex, v2.10

      4-20mA, version 02.10.zz. Push button on electronics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71026359

      Electronics HART Ex, ext. p. b., v2.10

      4-20mA, version 02.10.zz. Push button on housing. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71196092

      Electronic HART Ex, v2.11

      4-20mA, version 02.11.zz. Push button on electronics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 71196093

      Electronics HART Ex, ext. p. b., v2.11

      4-20mA, version 2.11.zz. Push button on housing. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 71263933

      Electronics HART Ex, v2.20

      4-20mA. Software version 02.20.zz, HART 7, DevRev22. Push button on electronics. Usage: FMB70, FMD77/FMD78, PMC71, PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 71263934

      Electronics HART Ex, ext. p. b., v2.20

      4-20mA. Software version 02.20.zz, HART 7, DevRev22. Push button on housing. Usage: FMB70, FMD77/FMD78, PMC71, PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 71386147

      Elektronics HART Ex, v2.30

      4-20mA. Software version 02.30.zz, HART 7, DevRev23. Push button on electronics. Usage: FMB70, FMD77/FMD78, PMC71, PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 71386153

      Electronics HART Ex, ext. p. b., v2.30

      4-20mA. Software version 02.30.zz, HART 7, DevRev23. Push button on housing. Usage: FMB70, FMD77/FMD78, PMC71, PMD75, PMP71/PMP75. Not for MID parts certificate (PMP71).

    • Nro 30 Tilauskoodi 52026762

      Upgrade electronics HART Ex, v2.1

      Consists of: electronics HART Ex, version 2.1. Push button on electronics. Display housing. Usage: conversion electronics 1.0 > 2.1, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 52026763

      Upgrade electr. HART Ex,ext.push b.,v2.1

      Consists of: electronics HART Ex, version 2.1. Push button external. Push button, housing HART version 2. Display housing. Usage: conversion electronics 1.0 > 2.1, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71071824

      Electronics PA, HW02.00, SW4.0

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 04.00.zz. Current consumption 13mA +/-1mA. Push button on electronics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71071825

      Electronics PA, ext. p.b., HW02.00,SW4.0

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 04.00.zz. Current consumption 13mA +/-1mA. Push button on housing. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71071826

      Electronics FF, HW2.00, SW3.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 03.00.zz. Current consumption 15mA +/- 1mA. Push button on electronics. Verwendung: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71071827

      Electronics FF, ext. p.b., HW2.00,SW3.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 03.00.zz. Current consumption 15mA +/- 1mA. Push button on housing. Verwendung: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71214057

      Electronics PA, Ex, HW2.00, SW3.00

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 03.00.zz. Current consumption 13mA +/-1mA. Push button on electronics. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71214064

      Electronics PA, ext. p.b.,HW2.00, SW3.00

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 03.00.zz. Current consumption 13mA +/-1mA. Push button on housing. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71214066

      Electronics FF, ext. p.b.,HW2.00, SW2.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 02.00.zz. Current consumption 15mA +/-1mA. Push button on housing. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71214069

      Electronics FF, HW2.00, SW2.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 02.00.zz. Current consumption 15mA +/-1mA. Push button on electronics. Usage: FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71214890

      Electronics FF, ext. p.b.,HW2.00, SW4.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 04.00.zz, FF, DevRev07. Current consumption 15mA +/-1mA. Push button on housing. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. Not for SIL.

    • Nro 30 Tilauskoodi 71214892

      Electronics FF, HW2.00, SW4.00

      FOUNDATION Fieldbus, Ex. Hardware version 02.00. Software version 04.00.zz, FF, DevRev07. Current consumption 15mA +/-1mA. Push button on electronics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75. Not for SIL.

    • Nro 30 Tilauskoodi 71224775

      Electronics PA, HW02.00, SW4.01

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 04.01.zz, PROFIBUS PA, Profile 3.02, DevRev03. Current consumption 13mA +/-1mA. Push button on electronics. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71224776

      Electronics PA, ext.p. b.,HW02.00,SW4.01

      PROFIBUS PA, Ex. Hardware version 02.00. Software version 04.01.zz, PROFIBUS PA, Profile 3.02, DevRev03. Current consumption 13mA +/-1mA. Push button on housing. Usage: FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP75.

    • Nro 30 Tilauskoodi 71327714

      Electronics 1-5VDC Low Power

      Usage: PMP71. Approval: V,T,X.

  • 31
  • 35
  • 40
    • Nro 40 Tilauskoodi 71002865

      Display module VU340, Cbar/Dbar/Dpilot S

      Digital display, 3 push buttons, snapped bracket. Languages DE/EN/FR/ES/IT/NL, from software version 02.00.zz; Usage: electronics HART. Languages DE/EN/FR/ES/IT/NL/JA/ZH. From software version 02.01.zz; Electronics HART. Languages DE/EN/FR/ES/IT/JA/ZH/RU, from software version 02.11.zz; Electronics HART. From software version 04.00.zz; Electronics PROFIBUS PA. From software version 03.00.zz; Electronics FOUNDATION Fieldbus, FMB70, FMD76/FMD77/FMD78, PMC71, PMD70/PMD75, PMP71/PMP72/PMP75. From software version 01.00.zz; Electronics HART, FMD71/FMD72.